A balance's display division d and its verification interval e serve different purposes. The first describes the indication resolution; the second is used in legal-metrology classification and assessment. Their values need not be identical.
The article explains why a test result must be compared with the correct acceptance basis. A tolerance expressed in verification intervals is different from one established by a laboratory's own measurement requirement.
Before evaluating repeatability, eccentricity or another performance characteristic, identify the applicable procedure, quantity and limit. Record them with the result. This avoids judging an instrument against a convenient display unit when the assessment actually requires a different criterion.
Source: RADWAG technical article.
Related technical articles:
Basic metrological tests I: installation and operating conditions
Basic metrological tests II: preparing and performing balance checks
Basic metrological tests III: minimum weight and analytical-balance checks