When to adjust a laboratory balance and how to plan calibration

on
Categories: Knowledge Base

Adjustment and calibration serve different purposes. Adjustment changes the instrument response; calibration establishes its measurement relationship and associated uncertainty. An automatic internal adjustment is therefore not a substitute for a complete calibration programme.

The source article highlights adjustment after installation or relocation and when temperature conditions change. Follow the instrument instructions for the applicable triggers and make sure the weighing pan is ready for the procedure.

Plan recalibration separately using the required uncertainty, instrument stability, use, environment and previous results. Intermediate checks help identify changes between calibrations. Use a documented schedule appropriate to the application rather than assuming one interval fits every laboratory balance.

Calibration intervals: OIML/ILAC guidance

Source: RADWAG technical article.