Automatic Device for Multichannel Pipette Calibration

Code: N/A

Compare products
Quote list
Views11

Compare configurations

SpecificationAP-12.5Y Automatic Device for Multichannel Pipette Calibration
WL-101-0416
AP-12.1.5Y Automatic Device for Multichannel Pipette Calibration
WL-101-0423
Maximum capacity [Max]52 g18 g
Readability [d]0,01 mg1 µg
Tare range-52 g-18 g
Standard repeatability [5% Max]5 µg2,8 µg
Standard repeatability [Max]10 µg3 µg for a single-channel balance 4 µg for a multichannel balance
Minimum weight (USP)10 mg2 mg
Minimum weight (U=1%, k=2)1 mg0,2 mg
Linearity±0,03 mg±0,01 mg
Stabilization time4 smax 10 s
Maximum volume12 ml4,5 ml
Device dimensions W x D x HNot available431x315,2x212 mm

 

AP-12.5Y

 
 

The AP-12.5Ycan be used for multichannel pipette tests with nominal volumes below 20 µl. The device features a readability [d] of 0.01 mg and can calibrate up to 12-channel fixed-volume and variable-volume pipettes.

 

Because the pipette calibration on the AP-12.5Y is fully automatic, it is four times faster. The calibration procedure for a 12-channel pipette takes less than 20 minutes, not six hours, on this instrument.

 

When purchasing the AP-12.5Y, you get:

  • two calibrated sensors: THB S and THB W,
  • two weighing pans: standard, and for balance test and calibration,
  • two pumps: a single-channel pump and a four-channel pump

 

The ergonomic design of the AP-12.5Y makes it possible to empty weighing vessels, and their overfilling is signalled by Ambient Light. When calibrating a multichannel pipette, a four-channel pump is used to simultaneously remove water from four vessels, greatly reducing the time required for pipette calibration.

 

After emptying a 12-channel pipette, the AP-12.5Y checks all channels in a single cycle. In addition to a database of pipette calibration reports, the AP-12.5Y includes a database of pipettes that is editable and contains maximum permissible error values according to ISO 8655. The report database, in turn, contains measurements and error values calculated for individual points and volumes. The user determines the format and content of the reports.

 

The AP-12.5Y monitors the ambient conditions using calibrated sensors: THB S (ambient conditions measurement) and THB W (liquid temperature measurement), in accordance with ISO 8655. To calibrate the device with mass standards, the weighing pan must be replaced with a pan dedicated to checking and calibrating the balance (both are included in the standard equipment).

 

The AP-12.5Y can calibrate not only 12-channel pipettes, but also single-channel pipettes. 

Maximum capacity [Max]52 g
Readability [d]0,01 mg
Tare range-52 g
Weighing pan dimensionsintended + ø26 mm for calibration
Operating temperature+10 – +40 °C
Relative humidity20% – 80%
Packaging dimensions W x D x H605×560×775 mm
Net weight17 kg
Gross weight23 kg
Power supplyAdapter: 100 – 240V AC 50/60Hz 1A Max; 15V DC 2,4A
Device: 12 – 15V DC 1,6A max; 10 – 29W**
Communication interface2×USB-A, USB-C, RS 232 (COM3), HDMI, Ethernet, Wi-Fi, Hotspot
Display10″ graphic colour touchscreen
Adjustmentinternal (automatic)
Standard repeatability [5% Max]5 µg
Standard repeatability [Max]10 µg
Minimum weight (USP)10 mg
Minimum weight (U=1%, k=2)1 mg
Linearity±0,03 mg
Stabilization time4 s
Operating temperature change rate±0,3 °C / 1 h (±1 °C / 8 h)
Relative humidity change rate±1% / h (±4% / 8 h)
Typical measurement time for a 12-channel pipette with constant volume, <br>according to ISO 8655 (for d=10 µg)*under 20 minutes
Typical measurement time for a 12-channel pipette with variable volume, <br>according to ISO 8655 (for d=10µg)*under 60 minutes
Maximum volume12 ml
  • Air density correction
  • ALIBI Memory
  • Ambient conditions monitoring
  • Animal weighing
  • Automatic sliding door
  • Autotest
  • BMI
  • Checkweighing
  • Density determination
  • Differential weighing
  • Dosing
  • Dry mass determination
  • Drying modes
  • Formulation
  • GLP Procedures
  • Internal battery
  • IR sensors
  • Labelling
  • Mass for titrator
  • Microwave drying
  • Moisture content analysis
  • Moveable range
  • Newton unit measurement
  • Packaged Goods Control
  • Parts counting
  • Peak hold
  • Percent Weighing
  • Pipettes Calibration
  • Plus/Minus Control
  • Replaceable unit
  • Samples drying
  • Statistical Quality Control
  • Statistics
  • Totalizing
  • Transactions
  • Under-pan weighing
  • Wi-Fi
  • Additional information line
  • Colour Touch Screen
  • Communicationinterfaces
  • Data exchange throughUSB storage devices
  • Electroniclevel indicator
  • Ergonomic Mechanical Design
  • Extra function keys
  • Grill pan
  • Infraredproximity sensors
  • Interfaces
  • Levelling System
  • Modes
  • Portability
  • SMARTLab
  • USB at the Balance Front
  • CE
  • 1 year warranty
AP-12.5Y – Speed, Process Automation, Cost Reduction Open video
AP-12 Automatic Calibration of Multichannel Pipettes in Your Hands Open video
ELLIPSIS – New Balance by RADWAG Open video
ELLIPSIS – Cutting Edge, Innovative, New Open video
Talk to it. Voice Commands in New ELLIPSIS Balance by RADWAG Open video
ELLIPSIS – Scan or Tap to Log In Open video
Remote VPN Support on ELLIPSIS Balances Open video
ELLIPSIS – Unlimited Connections Open video
Video 9 Open video
Video 10 Open video
Video 11 Open video
Video 12 Open video
Video 13 Open video