Antistatic ionizer is used to remove overflow of mainly plus ions, and also minus ions from the load and elements of a balance and the environment. Electric charges of the air are neutralized by the ionizer enabling reaching a quasi-balance of plus and minus ions.
Unbalance of plus and minus ions in the air frequently causes unwanted electric influences (attraction or repulsion) of elements which are in small distance to each other. It is a usual phenomenon for non-conducting materials (i.e. plastic, glass). As some of balance elements are made of plastic and glass, this problem can considerably affect their operation and measurement results. The influence of electrostatics appears in a form of multiplied repeatability error or as multiplied error of indications.
DJ-04
Antistatic ionizer is used to remove overflow of mainly plus ions, and also minus ions from the load and elements of a balance and the environment. Electric charges of the air are neutralized by the ionizer enabling reaching a quasi-balance of plus and minus ions.
Unbalance of plus and minus ions in the air frequently causes unwanted electric influences (attraction or repulsion) of elements which are in small distance to each other. It is a usual phenomenon for non-conducting materials (i.e. plastic, glass). As some of balance elements are made of plastic and glass, this problem can considerably affect their operation and measurement results. The influence of electrostatics appears in a form of multiplied repeatability error or as multiplied error of indications.
Select a configuration to view its specifications.
Operating temperature
0 – +50 °C
Relative humidity
5% – 80% no condensation
Packaging dimensions W x D x H
375×375×150 mm
Net weight
1,3 kg
Gross weight
2,5 kg
Power supply
power supply: 100 – 240V AC 50/60Hz 1A Max; 15V DC 2,4A ionizer: 12-15V DC 2W
Distance form a sample
5 – 200 mm (recommended 5 – 50 mm)
Frame dimensions
175×70×172 mm
High voltage
5 kV DC, pulsed, peak value
Maximum short-circuit current
<2mA (output HV)
Neutralization method
Corona discharge, alternating emission of positive and negative ions
Neutralization time
<5 s (typically <2 s)
Operating temperature
0 – +50 °C
Relative humidity
5% – 80% no condensation
Packaging dimensions W x D x H
375×375×150 mm
Net weight
1,3 kg
Gross weight
2,5 kg
Power supply
power supply: 100 – 240V AC 50/60Hz 1A Max; 15V DC 2,4A ionizer: 12-15V DC 2W
Distance form a sample
5 – 200 mm (recommended 5 – 50 mm)
Frame dimensions
268×110×295 mm
High voltage
5 kV DC, pulsed, peak value
Maximum short-circuit current
<2mA (output HV)
Neutralization method
Corona discharge, alternating emission of positive and negative ions
Neutralization time
<5 s (typically <2 s)
Select a configuration to view its manufacturer information.