Select a configuration to view its description.
SDKP
Application:
Calibration of automatic pipettes:
- with fixed volume
- with adjustable volume.
Unit contents:
- table for calibration
- anti-vibration table
- probe for temperature, humidity and pressure measurement
- probe for water temperature measurement
- sink for used water
- armrests
- LCD monitor
- PC with pre-installed software PIPETTES RADWAG
- power supply
- keyboard and mouse
- container for used water
Adapters for pipettes calibration
Pipette calibration adapters are intended for calibration and verification of piston pipettes using gravimetric method. Use of the pipette calibration adapter reduces maximum capacity of the balance by the mass of the weighing vessel. The table presents the minimum reguirements in accordance with ISO 8655 standard. It also enables to select suitable adapter for pipette calibration and respective balance version.
|
|
Resolution mg |
Repeatability and linearity mg |
Recommended balance | ||||||||||
| 1 µl ≤ V ≤ 10 µl | 0,001 | 0,002 |
MYA 21.5Y PLUS
MYA 21.5Y
|
XA 6/21.5Y.M.A PLUS
XA 6/21.5Y.M.A
XA 6/21.5Y.M
|
|||||||||
| 10 µl ≤ V ≤ 100 µl | 0,01 | 0,02 |
XA 82/220.5Y.A PLUS
XA 82/220.5Y.A
XA 82/220.5Y
|
XA 120/250.5Y.A PLUS
XA 120/250.5Y.A
XA 120/250.5Y
|
|||||||||
| 100 µl ≤ V ≤ 1000 µl | 0,1 | 0,2 | |||||||||||
| 1 ml < V ≤ 10 ml | 0,1 | 0,2 | |||||||||||
| Adapter for pipette calibration | MY 11 | XA 11 | XA 17 | XA 100 | XA 17 | XA 100 | |||||||
| Maximum vessel capacity | 11 ml | 11 ml | 17 ml | 100 ml | 17 ml | 100 ml | |||||||
| Weighing vessel mass | ~ 10 g | ~ 10 g | ~ 11 g | ~ 65 g | ~ 11 g | ~ 65 g | |||||||
1 For practical purposes, the nominal volume may be used to choose the balance [Reference: ISO8655-6:2002, pg. 2].
If the standard uncertainty of measurement of the balance is known (e.g. from the balance calibration certificate), it may be used instead of the repeatability and linearity. The standard uncertainty of measurement shall not be more than two to three times the resolution [Reference: ISO8655-6:2002, pg. 2].
a{color:#737373;}
SDKPDUAL
Application:
Calibration of automatic pipettes:
- with fixed volume
- with adjustable volume.
Unit contents:
- table for calibration
- anti-vibration table x2
- probe for temperature, humidity and pressure measurement
- probe for water temperature measurement
- sink for used water
- armrests
- LCD monitor
- PC with pre-installed software PIPETTES RADWAG
- absorbtion pump
- power supply
- keyboard and mouse
- container for used water
Adapters for pipettes calibration
Pipette calibration adapters are intended for calibration and verification of piston pipettes using gravimetric method. Use of the pipette calibration adapter reduces maximum capacity of the balance by the mass of the weighing vessel. The table presents the minimum reguirements in accordance with ISO 8655 standard. It also enables to select suitable adapter for pipette calibration and respective balance version.
|
|
Resolution mg |
Repeatability and linearity mg |
Recommended balance | ||||||||||
| 1 µl ≤ V ≤ 10 µl | 0,001 | 0,002 |
MYA 21.5Y PLUS
MYA 21.5Y
|
XA 6/21.5Y.M.A PLUS
XA 6/21.5Y.M.A
XA 6/21.5Y.M
|
|||||||||
| 10 µl ≤ V ≤ 100 µl | 0,01 | 0,02 |
XA 82/220.5Y.A PLUS
XA 82/220.5Y.A
XA 82/220.5Y
|
XA 120/250.5Y.A PLUS
XA 120/250.5Y.A
XA 120/250.5Y
|
|||||||||
| 100 µl ≤ V ≤ 1000 µl | 0,1 | 0,2 | |||||||||||
| 1 ml < V ≤ 10 ml | 0,1 | 0,2 | |||||||||||
| Adapter for pipette calibration | MY 11 | XA 11 | XA 17 | XA 100 | XA 17 | XA 100 | |||||||
| Maximum vessel capacity | 11 ml | 11 ml | 17 ml | 100 ml | 17 ml | 100 ml | |||||||
| Weighing vessel mass | ~ 10 g | ~ 10 g | ~ 11 g | ~ 65 g | ~ 11 g | ~ 65 g | |||||||
1 For practical purposes, the nominal volume may be used to choose the balance [Reference: ISO8655-6:2002, pg. 2].
If the standard uncertainty of measurement of the balance is known (e.g. from the balance calibration certificate), it may be used instead of the repeatability and linearity. The standard uncertainty of measurement shall not be more than two to three times the resolution [Reference: ISO8655-6:2002, pg. 2].
a{color:#737373;}
PRODUCT DESCRIPTION - WORKSTATION FOR PIPETTES CALIBRATION XA 52
Application:
1. Calibration of automatic pipettes from 1µl to 10ml:
- with fixed volume
- with adjustable volume.
Unit contents:
- table for calibration
- XA 52.4Y analytical balance
- anti-vibration table
- probe for temperature, humidity and pressure measurement
- probe for water temperature measurement
- sink for used water
- armrests
- LCD monitor
- PC with pre-installed software PIPETTES RADWAG
- power supply
- keyboard and mouse
- container for used water
* The pipettes and other small accessories are not included in the unit content.
PRODUCT DESCRIPTION - WORKSTATION FOR PIPETTES CALIBRATION XA 82
Application:
1. Calibration of automatic pipettes from 1µl to 10ml:
- with fixed volume
- with adjustable volume.
Unit contents:
- table for calibration
- XA 82/220.4Y analytical balance
- anti-vibration table
- probe for temperature, humidity and pressure measurement
- probe for water temperature measurement
- sink for used water
- armrests
- LCD monitor
- PC with pre-installed software PIPETTES RADWAG
- power supply
- keyboard and mouse
- container for used water
* The pipettes and other small accessories are not included in the unit content.
Nominal value - 1 g.