Moisture determination in bath and shower gel: sample preparation and validation

on
Categories: Application Notes, Knowledge Base

This application note compares infrared moisture analysis of bath and shower gel with an oven-drying reference procedure. The supporting English document provides the tested preparation, measurement settings and experimental results.

Mix the gel before collecting a portion and keep the material in closed packaging before the test. The reference arrangement uses sand and a glass rod to distribute the sample and limit skin formation.

Use the document to select the sample quantity, heating conditions and endpoint for the particular material. Compare repeated results with the stated reference and assess both agreement and variability. The published settings describe the tested samples; establish their suitability for local products and the required measurement performance before using them as a routine method.