The BelGIM case describes creating a liquid-density reference system using a calibrated silicon sphere, hydrostatic weighing and temperature control. The sphere was weighed in air and in the liquid under investigation with an AGV4-1000 comparator.
Reference hydrometers were then weighed in air and in the same liquid using an XA 210.4Y balance. This linked the density values to the hydrometers through a controlled measurement process. A circulating bath and climatic chamber supported stable conditions.
The reported system covered liquid densities from 650 to 2,000 kg/m³. The practical lesson is the relationship between traceability, the comparison sequence and environmental control; the comparator specification alone does not describe the complete reference method.