How cutting particulate-matter filters can affect gravimetric analysis

on
Categories: Knowledge Base

A study reported by RADWAG examined material losses when PM-loaded filters were cut for further analysis. The work used gravimetric measurements alongside Karl Fischer titration and ion chromatography on PM1 and PM2.5 samples collected in Zabrze.

The reported mean cutting losses were approximately 0.060 mg for PM1 filters and 0.192 mg for PM2.5 filters. Relative to the deposited material, the losses were about 3.3–3.5% and 6–10%, respectively, in the studied samples. These are experimental results, not general correction factors.

The study also identified moisture as an important influence on measurement accuracy. When dividing filters, account for handling losses and conditioning, and evaluate how the preparation method affects the quantity subsequently reported. The source links to the Chemosphere paper by Chyzhykov, Widziewicz-Rzońca and Mathews.

Original research paper on PM filter cutting

Source: RADWAG technical article.

Views6